Page #158 background image
11-6
To disassemble the expiration valve assembly:
1. Rotate the expiration valve handwheel counter-clockwise until the indicating arrow
on the handwheel aligns with the
position. Then pull out the expiration valve
assembly horizontally.
2. Remove the expiration valve membrane.
To install the expiration valve assembly:
1. Install the expiration valve membrane onto the expiration valve assembly.
2. Ensure the indicating arrow
on the handwheel aligns with the position. Push
the expiration valve assembly into the corresponding connector on the ventilator
horizontally to the end. Then rotate the expiration valve handwheel clockwise (and
depress the handwheel in the direction the expiration valve is installed) until the
indicating arrow
on the handwheel aligns with the position.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray SV300 and is the answer not in the manual?

Mindray SV300 Specifications

General IconGeneral
Waveform DisplayPressure, Flow, Volume
Battery BackupUp to 2 hours
Ventilation ModesVCV, PCV, SIMV
Tidal Volume Range20 to 2000 mL
FiO221 to 100%
Inspiratory Time0.2-5.0 s
Monitoring ParametersPaw, Ppeak, Pmean, Pplat, PEEP, VTI, VTE, MV, f, FiO2, I:E
AlarmsHigh/Low Airway Pressure, High/Low Minute Volume, High/Low Tidal Volume, High/Low Respiratory Rate, High/Low FiO2, Apnea, Power Failure
Power SupplyAC 100-240V, 50/60Hz
Peak Flow180 L/min
Trigger Sensitivity (Flow)0.5-10 L/min