11-11
To disassemble the HEPA filter assembly and air intake dust filter:
1. Pull the two snaps on the main unit air inlet grilleto remove the grille.
2. Pull the snap on the HEPA filter to take it out. If it is necessary to remove the air intake
dust filter, pinch the dust filter with two fingers and take it out.
To install the HEPA filter assembly and air intake dust filter:
1. Align the HEPA filter with the corresponding slot, and push in the direction the HEPA
filter is installed.
2. Fasten the snap of the HEPA filter.
3. Check the snap on the HEPA filter and make sure it is fastened in place.
4. Install the main unit air inlet grille.
NOTE
Install the specified HEPA filter and air intake dust filter.
CAUTION
Do not run the ventilator if the ventilator is not equipped with HEPA filter to avoid
contaminating the ventilator inspiration port and patient tubing.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray SV300 and is the answer not in the manual?

Mindray SV300 Specifications

General IconGeneral
Waveform DisplayPressure, Flow, Volume
Battery BackupUp to 2 hours
Ventilation ModesVCV, PCV, SIMV
Tidal Volume Range20 to 2000 mL
FiO221 to 100%
Inspiratory Time0.2-5.0 s
Monitoring ParametersPaw, Ppeak, Pmean, Pplat, PEEP, VTI, VTE, MV, f, FiO2, I:E
AlarmsHigh/Low Airway Pressure, High/Low Minute Volume, High/Low Tidal Volume, High/Low Respiratory Rate, High/Low FiO2, Apnea, Power Failure
Power SupplyAC 100-240V, 50/60Hz
Peak Flow180 L/min
Trigger Sensitivity (Flow)0.5-10 L/min