11-15
To disassemble the patient tubing:
Pull out the patient tubing one by one.
To install the patient tubing:
1. Mount the filters onto the inspiratory and expiratory ports.
2. Connect the inspiratory filter to the water trap via the tube. Connect the other end of the
tube to the Y piece.
3. Connect the expiratory filter to the water trap via the tube. Connect the other end of the
tube to the Y piece.
4. Connect the patient side of the Y piece to the HME and then connect the patient to the
HME.
5. Place the patient tubing onto the support arm hook.