B-9 
Control accuracy (O
2
 Therapy) 
Continuous Flow  ± (2 L/min+10 % of setting) (BTPS) 
O
2
 Concentration  ± (3 Vol.% +1 % of setting) 
Monitoring accuracy 
Ppeak 
± (2 cmH
2
O + 4 % of the actual reading) 
Pplat 
Pmean 
PEEP 
TVi 
0 mL~100 mL: ± (10 mL + 3 % of the actual reading) (BTPS); 
100 mL~4000 mL: ± (3 mL + 10 % of the actual reading) (BTPS) 
TVe 
TVe/IBW 
TVe spn 
MV 
± (0.2 L/min + 10 % of the actual reading) (BTPS) 
MVspn 
MVleak 
ftotal 
±5 % of reading or ±1 bpm, whichever is greater 
fmand 
fspn 
Rinsp 
0 to 20: ±10 cmH
2
O/(L/s) 
Other range: 50%. 
Rexp 
Cstat 
± (2 mL/cmH
2
O + 20 % of the actual reading) 
Cdyn 
RSBI  ± (3 1/(L•min)+15 % of the actual reading) 
WOB  ± (1 J/min+15 % of the actual reading) 
NIF  ± (2 cmH
2
O + 4 % of the actual reading) 
P0.1  ± (2 cmH
2
O + 4 % of the actual reading) 
PEEPi No declaration 
RCexp  ± (0.2 s + 20 % of the actual reading) 
FiO
2
  ± (2.5 Vol. % + 2.5 % of the actual reading) 
Tinsp  ± 0.05 s 
I:E  ±50 ms or ±6%, whichever is greater. 
Oxygen concentration 
controlling response time 
the responded time of the oxygen concentration in the delivered 
volume to change from a volume fraction of 21 % to 90 % of the 
maximum settable oxygen concentration: 
when TV=500 mL, f=10 bpm, I:E=1:2,  ≤90 s ; 
when TV=150 mL, f=20 bpm, I:E=1:2,  ≤120 s; 
when TV=30 mL, f=30 bpm, I:E=1:2,  ≤90 s.