B-9
Control accuracy (O
2
Therapy)
Continuous Flow ± (2 L/min+10 % of setting) (BTPS)
O
2
Concentration ± (3 Vol.% +1 % of setting)
Monitoring accuracy
Ppeak
± (2 cmH
2
O + 4 % of the actual reading)
Pplat
Pmean
PEEP
TVi
0 mL100 mL: ± (10 mL + 3 % of the actual reading) (BTPS);
100 mL4000 mL: ± (3 mL + 10 % of the actual reading) (BTPS)
TVe
TVe/IBW
TVe spn
MV
± (0.2 L/min + 10 % of the actual reading) (BTPS)
MVspn
MVleak
ftotal
±5 % of reading or ±1 bpm, whichever is greater
fmand
fspn
Rinsp
0 to 20: ±10 cmH
2
O/(L/s)
Other range: 50%.
Rexp
Cstat
± (2 mL/cmH
2
O + 20 % of the actual reading)
Cdyn
RSBI ± (3 1/(L•min)+15 % of the actual reading)
WOB ± (1 J/min+15 % of the actual reading)
NIF ± (2 cmH
2
O + 4 % of the actual reading)
P0.1 ± (2 cmH
2
O + 4 % of the actual reading)
PEEPi No declaration
RCexp ± (0.2 s + 20 % of the actual reading)
FiO
2
± (2.5 Vol. % + 2.5 % of the actual reading)
Tinsp ± 0.05 s
I:E ±50 ms or ±6%, whichever is greater.
Oxygen concentration
controlling response time
the responded time of the oxygen concentration in the delivered
volume to change from a volume fraction of 21 % to 90 % of the
maximum settable oxygen concentration:
when TV=500 mL, f=10 bpm, I:E=1:2, 90 s ;
when TV=150 mL, f=20 bpm, I:E=1:2, 120 s;
when TV=30 mL, f=30 bpm, I:E=1:2, 90 s.