B-12
Function Specification
NIF Maximum negative pressure produced by patient’s spontaneous breathing within
a period of time.
PEEPi The PEEPi measure function supports measurement of two parameters: PEEPi
and Vtrap. PEEPi is the positive end-expiratory pressure produced by the
trapped gas and Vtrap is the trapped gas volume.
P-V By drawing static pressure-volume loop (static P-V loop), P-V tool is the method
to determine the optimal PEEP based on the characteristic points on the static
P-V loop.
ATRC ATRC stands for the function of automatic tube resistance compensation. By
selecting appropriate endotracheal (ET) tube or tracheostomy (Trach) tube of
different diameters for the user, the ventilator can adjust gas delivery pressure
automatically.
Sigh The sigh function is used to open collapsed areas of the lung or to keep the lung
open.
The sigh function can be activated in all ventilation modes except CPAP/PSV,
DuoLevel, and APRV.
Each time after the sigh function is activated, ventilation is controlled based on
the user-set sigh ventilation cycles and the set value of int.PEEP. PEEP of the
sigh ventilation cycle increases int.PEEP level. After that, sigh is
automatically switched off until next sigh time interval.
Screen Locking Prevents ventilator settings and values displayed from being changed due to
inadvertent key clicking.
O
2
Therapy Continuous flow application with adjustable O
2
concentration and flow for
patients with independent breathing and using oxygen masks.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray SV300 and is the answer not in the manual?

Mindray SV300 Specifications

General IconGeneral
Waveform DisplayPressure, Flow, Volume
Battery BackupUp to 2 hours
Ventilation ModesVCV, PCV, SIMV
Tidal Volume Range20 to 2000 mL
FiO221 to 100%
Inspiratory Time0.2-5.0 s
Monitoring ParametersPaw, Ppeak, Pmean, Pplat, PEEP, VTI, VTE, MV, f, FiO2, I:E
AlarmsHigh/Low Airway Pressure, High/Low Minute Volume, High/Low Tidal Volume, High/Low Respiratory Rate, High/Low FiO2, Apnea, Power Failure
Power SupplyAC 100-240V, 50/60Hz
Peak Flow180 L/min
Trigger Sensitivity (Flow)0.5-10 L/min