B-17
Effect of interfering gas
Gas under test
Error (O
2
)
50% He/50% O
2
<1 %
80% N
2
O/20% O
2
1 % to 1.5 %
4% Halothane/28.8% O
2
/67.2% N
2
O 1.5 % to 2 %
5% Sevoflurane/28.5% O
2
/ 66.5%
N
2
O
1 % to 1.5 %
5% Enflurane/28.5% O
2
/66.5% N
2
O 1.2 % to 1.8 %
5% Isoflurane/28.5% O
2
/66.5% N
2
O 1.2 % to 1.8 %
5% CO
2
/ 28.5% O
2
/66.5% N
2
O <1 %
Theory of Operation
O
2
sensor can monitor the patient’s FiO
2
. O
2
sensor is of the self-powered, diffusion limited,
metal-air battery type comprising an anode, electrolyte, diffusion barrier and air cathode as shown
below:
At the cathode oxygen is reduced to hydroxyl ions according to the equation:
OHeOHO 442
22
The hydroxyl ions in turn oxidise the metal anode as follows:
eOHPbOOHPb 42242
2
Overall the cell reaction may be represented as:
Air supply
Cathode
Electrolyte
Anode
Load resistor
Solid membrane
Diffusion barrier

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray SV300 and is the answer not in the manual?

Mindray SV300 Specifications

General IconGeneral
Waveform DisplayPressure, Flow, Volume
Battery BackupUp to 2 hours
Ventilation ModesVCV, PCV, SIMV
Tidal Volume Range20 to 2000 mL
FiO221 to 100%
Inspiratory Time0.2-5.0 s
Monitoring ParametersPaw, Ppeak, Pmean, Pplat, PEEP, VTI, VTE, MV, f, FiO2, I:E
AlarmsHigh/Low Airway Pressure, High/Low Minute Volume, High/Low Tidal Volume, High/Low Respiratory Rate, High/Low FiO2, Apnea, Power Failure
Power SupplyAC 100-240V, 50/60Hz
Peak Flow180 L/min
Trigger Sensitivity (Flow)0.5-10 L/min