4-34 
 
10.  Verify that the tidal volume displayed on the anesthesia machine is within 7% (±42 mL) 
of the set value within approximately the first 1 minute after ventilation starts. 
11.  Verify that the measured O2 concentration is at least 97% after 5 minutes 
12.  Set the air flow to 3 L/min and disable the O2 flow rates. 
13.  Verify that the measured O2 concentration is 21% ±3% (volume fraction) after 5 
minutes. 
 
4.13.4 VCV Child Ventilation Mode Test 
1.  Attach a breathing circuit and breathing bag. 
NOTE 
  For testing purposes always use a reusable breathing circuit. 
2.  Attach an adult test lung to the Y-piece of the breathing circuit. 
NOTE 
  Limit the volume in the test lung to provide sufficient airway pressure to satisfy the 
Low Peak Pressure alarm. Or set the peak pressure alarm limit to a smaller value 
to prevent the alarm when an adult test lung is used. 
3.  Attach a ventilation tester between the expiratory port and the breathing hose. 
4.  Select Air in [Balance Gas] and single-tube flow in [Fresh Gas Control]. Set the O2 flow 
to 2 L/min and set the N2O and air flow rates to the minimum value. 
5.  Set breathing parameters under standby mode. 
Parameter  Setting 
Patient Size 
Child 
Ventilation Mode 
VCV 
Vt 120 
Rate 15 
I:E 1:2 
Tpause 10 
PEEP Off 
Plimit 40 
6.  Press the [Set Mode] button to exit standby mode and begin ventilation. 
7.  Verify that the pressure waveform, respiratory capacity, mean or plateau pressure, 
respiratory rate, and minute ventilation appear on the screen. 
8.  Verify that the displayed tidal volume is within 15 mL of the set value within 
approximately the first 1 minute after ventilation starts. 
9.  Verify that the tidal volume, measured by a ventilation tester at the expiratory port, is 
within 15 mL of the set value within approximately the first 1 minute after ventilation 
starts. 
 
4.13.5 Air Way Disconnection Alarm Test 
1.  Disconnect the expiratory limb from the expiratory port of breathing system when the 
ventilator is working. 
2. Ensure the following alarms appear: 
  A low Paw alarm appears on the screen 
  Alarm sounds