2-64 
 
2.2.10 Drive Pressure - High Pressure Regulator (200 kPa, 29 
psi) 
The drive pressure regulator stabilizes the supply pressure upon the proportional valve. 
The flow generated by the proportional valve is therefore independent of pressure variations at the 
supply. 
Setting the drive pressure regulator at 200 kPa (29 psi) allows for a maximum of inspiratory flow 
of 110 L/min at the ventilator. 
 
 
2.2.11 Drive Gas Assembly 
The drive gas assembly comprises the inspiratory circuit and PEEP circuit. The gas flow of the 
inspiratory circuit goes through the inspiratory valve, which generates a gas flow of 0 to 110 
L/min. The gas flow of the PEEP circuit goes through the PEEP valve, which also generates gas 
pressure of 3 to 30 cm H2O. Both the PEEP and inspiratory valves are proportional solenoid 
valves controlled by the valve drive board. 
 
 
2.2.12 Negative Pressure Suction System 
 
   
 
This negative pressure suction system is a Venturi negative pressure suction system. Negative 
pressure is produced by the Venturi pipe in the negative pressure generator drive d by Air. The 
negative pressure value is displayed and adjusted as needed by the negative pressure control pamel 
assembly. Drive gas goes through the negative pressure generator, and then is discharged from the 
muffler assembly to the air. Negative pressure control pamel realizes the on-off controlling and 
adjusting of the negative pressure source.   
 
 
Negative pressure 
generator 
Muffler 
Drive gas 
inlet 
Negative pressure 
outlet 
Negative 
pressure 
selecting 
knob 
Negative 
pressure 
adjusting knob 
Negative 
pressure gauge