8-78 
 
8.5.6 Diagnosing Pressure Data Exceptions of Sensors 
  The pressure sensor has one sampling line. Connection exceptions include: 
  The sampling line is not connected. 
  The sampling line is connected incorrectly. 
 
You can check whether the sampling lines are connected correctly by using the valve diagnosis 
tool. 
  Perform the following operations to check the connection of the sampling line of the 
pressure sensor: 
  During normal ventilation, sampling line connection errors will be easily noticed 
through the Paw waveform and technical alarms. If the waveform value decreases 
as the actual pressure increases and the alarm Paw Too Low or Patient Circuit Leak 
is generated, the sampling line of the airway pressure sensor may be connected 
incorrectly.  
  Access the Diagnostic Tests menu and set the PEEP safety valve to the ON state. 
Increase the D/A value of the PEEP valve gradually and observe whether the 
sampling A/D value measured by the PEEP pressure sensor increases accordingly. 
If no, it further proves that the PEEP pressure sensor may be connected incorrectly. 
  To diagnose whether the sampling line of the pressure sensor is connected normally in 
case of pressure calibration failure, perform the following steps: 
1.  Connect the tubes of the anaesthesia machine according to section 5.3Valves Diagnostic 
Test Tools.   
2.  Ensure that the water collection cup is reinstalled.   
3.  Ensure that the gas source complies with the regulations. Access the Diagnostic Tests 
menu and set the PEEP safety valve to the ON state. Set the flow volume of the 
inspiratory valve to 5 L/min.   
4.  Increase the D/A value of the PEEP valve gradually. The sampling A/D value measured 
by the pressure sensor increases accordingly.   
  If the sampling A/D value measured by a pressure sensor decreases gradually, the 
sampling line of the pressure sensor may be connected incorrectly. 
  If the sampling A/D value measured by a pressure sensor remains unchanged, the 
sampling line of the pressure sensor may not be connected. 
  The pressure value on the airway pressure gauge increases. Otherwise, the airway 
pressure gauge is damaged.