9-18
9.2.18 Remove EFCS Flowmeter Board
1. Remove the EFCS flowmeter.
2. Unscrew the four (4) screws fixing the EFCS flowmeter assembly and remove the
connected cables to take out the EFCS flowmeter assembly.
3. Unscrew the four (4) screws fixing the EFCS flowmeter board to take out the EFCS
flowmeter board.