8-37
8.4.1.1 Precautions for Use of Test Fixture of Pressure Sampling Tube of
Flow Sensor
The test fixture of the flow sensor tributary has two pressure sampling tube connectors, as shown
in the following figure.
The pressure sampling tube connector can be connected to the Silicon hose (6x300). When using
the test fixture of the flow sensor tributary, take out the expiration or inspiration flow sensor from
the patient circuit. Install the test fixture of the flow sensor tributary in the original position of the
expiration or inspiration flow sensor, and screw the cap of the breathing interface, as shown in the
following figure. Perform the tests after connecting the Silicon hose (6x300) to the pressure
sampling tube connector of the corresponding test tributary.
Pressure
sampling tube