2-19
Pin Name Function
6)
NTC2 Thermistor inside battery
7)
BC2 Battery in-place signal
8)
GND Ground
2.1.4.3 Auxiliary Electrical Outlet
Auxiliary electrical outlet comes from total AC mains inlet, and it is to supply the external devices
of anesthesia machine.
No. Sales area
Auxiliary
electrical
outlet
Declaration
of AC mains
inlet
Declaration of
total current of
auxiliary
electrical outlet
Declaration of
singal auxiliary
electrical outlet
1)
Domastic standard 4
220-240V~
6A 50/60Hz
5A MAX
220-240V~ 3A
50/60Hz
2)
American standard 4
100-120V~
7A 50/60Hz
5A MAX
100-120V~ 3A
50/60Hz
3)
European
standard/British
standard/Indian
standard/Australian
standard/ B
specification/Brazili
an standard
4
100-240V~
7A 50/60Hz
5A MAX
100-240V~ 3A
50/60Hz
4)
South African
standard
3
100-240V~
7A 50/60Hz
5A MAX
100-240V~ 3A
50/60Hz
5)
without auxiliary
electrical outlet
configuration
None
100-240V~
7A 50/60Hz
None None