2-52 
 
2.2.4.2 Electronic Flow Control System-EFCS 
 
 
 
 
 
 
The EFCS uses solenoid proportional valves to control the oxygen, air and nitrous oxide flow and 
monitor the actual flow, and provides feedback to the flow control unit. After passing three 
independent check valves, the gases of the three limbs are mixed at the rear end of the assembly 
and output. The O2-N2O ratio is fully controlled by a software algorithm. The system display 
outputs the monitored values of parameters such as the limb flow and O2-N2O ratio. 
The above figure shows the pneumatic block structure of the EFCS. The three gases, after being 
regulated, goes through the proportional valves, through limb sensors, through the gas mixing 
chamber, and into the three-way valve. Then, the gas mixture returns to the total flow sensor and 
finally output to the mechanical float flow meter.   
O2 (from the 
O2 supply) 
N2O (from 
the N2O 
su
l
 
Air (from the 
air supply) 
From 
backup O2 
flowmete
 
To the 
vaporizer 
manifold 
Balance gas 
sensor 
O2 sensor 
Total flow 
sensor