8-38 
 
 
8.4.1.2 Precautions for Use of Test Fixture of Vaporizer Support 
Before use, take out the seal ring on the connector of the vaporizer support for cooperating with 
the vaporizer. Slide the fixture into the connector, as shown in the following figures. 
 
 
Rotate the pressure head of the test fixture of the vaporizer support clockwise to the end until the 
lower surface of the test fixture of the vaporizer support is in contact with the upper surface of the 
vaporizer support connector, as shown in the following figure.   
After the test fixture slid in Before the test 
fixture slid in