3-18
3.2.3.1 Cylinder Wrench and User Guide
1. Fix the cylinder wrench onto the back of this device so that it can be used to open or
close cylinders without disconnecting from the machine.
2. Hang the user guide on the device handle.
3.2.4 Vaporizer
WARNING
If the vaporizer is incompatible with this device Anesthesia System, the vaporizer
will not work at all. Use vaporizers with Selectatec mounting system that are
compliant to ISO8835-4. Refer to the vaporizer manufacturer’s Instructions For
Use for filling or draining the vaporizer and other information.
For this device Anesthesia System, using or turning on more than one vaporizer
simultaneously is mechanically prevented by the Selectatec mount system. Do not
attempt to override this safety feature.
Use care in lifting and manipulating vaporizers during the mounting process as
their weight may be greater than expected, based on their size and shape.
NOTE
The barometric pressure at the customer side may be different from the factory
calibration pressure of anesthetic vaporizer. This may cause inaccurate
concentration output of anesthetic agent. The operator should continuously
monitor the concentration of anesthetic agent during system use.