3-20
4. When reinstalling the vaporizer, lift each vaporizer straight up off the manifold rather
than pulling forward. Do not rotate the vaporizer on the manifold.
5. If a vaporizer unintentionally lifts off the manifold, install it again and complete steps 1
through 3. If the vaporizer lifts off a second time, do not use the system.
NOTE
A Desflurane vaporizer may be mounted similarly as other vaporizers, but may
require a power cord. For more detailed instructions on installation and proper
use, refer to the specific manufacturer’s Instructions for Use of the Desflurane
vaporizer.
3.2.4.2 Fill the Vaporizer
NOTE
This device should use vaporizers with Selectatec mounting system that are
compliant to ISO8835-4. Refer to the vaporizer manufacturer’s Instructions For
Use for filling or draining the vaporizer and other information.
WARNING
Ensure that the correct anesthetic agent is used. The vaporizer is designed with the
specific anesthetic agent name on it and further indicated by color coded label. The
concentration of the anesthetic agent actually outputted will vary if the vaporizer is
filled with the wrong agent.
3.2.4.3 Drain the Vaporizer
WARNING
Do not reuse the agent drained from the vaporizer. Treat as a hazardous chemical
and follow local regulations for proper disposal.