EasyManua.ls Logo

Mindray WATO EX-55Pro - Power System; Power Board

Mindray WATO EX-55Pro
474 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
2-15
Pin No. Signal Name Discription
79) LCDVGA4+ LVDS clock signal
80) GND Ground
81) RXD_VCM1
Received signal of monitoring signal detection board
series port
82) TXD_VCM1
Sent signal of monitoring signal detection board series
port
83) GND Ground
84) RXD_VPM1 Received signal of auxiliary control module series port
85) TXD_VPM1 Sent signal of auxiliary control module series port
86) GND Ground
87) RXD_KEY Reserved series port signal
88) TXD_KEY Reserved series port signal
89) GND Ground
90) Touchpad_CON TouchPad power supply control signal
91) MAIN_ACGO_CTRL1 ACGO valve control signal 1
92) FAN_PWM1 Fan control signal
93) FAN_PWM2 Fan control signal
94) IN1 Backup input signal
95) NC No connection inside
96) NC No connection inside
2.1.4 Power System
After the total AC mains inlet enters the anesthesia machine, it will be divided into two. One is
supplied to power board to transfer into the AC supply of the anesthesia machine; the other is
supplied to auxiliary output to supply the external devices of anesthesia machine.
2.1.4.1 Power Board
AC area

Table of Contents

Related product manuals