7-6
7.3.1.2 Calibration (User) Procedure
Perform the following steps to calibrate flow sensors.
1. Enter the standby mode.
2. [Main] > [General]> [Calibrate Flow Sensors], to enter the screen shown below.
3. Follow the instructions on the screen to set the machine and select [Next] to enter the
screen shown below.