7-7
4. Select [Begin] to calibrate the flow sensor. During the calibration, you can select
[Cancel] to abort.
5. The screen shown below is displayed if the ongoing calibration is aborted. Select [Try
Again] to do the calibration again or select [Done] to exit the calibration screen.
6. The screen shown below is displayed if the flow sensor calibration fails. Select [Try
Again] to do the calibration again or select [Done] to exit the calibration screen.