7-15
interface. Press [ZERO] on the panel. In the displayed interface, press [OK] to complete zeroing.
(2) On the panel of the calibration device VT, press [MODE]. Select [Calibration Mode] from
the menu, and press [OK] on the panel to go to the calibration interface. See the following figure:
If the VT Plus is used as the calibration device, configure the Fluke VT Plus in the following way:
(1) Gas settings: Select [Main], and select [Setting] > [ENTER] > [Gas Settings] >
[MODIFY] > [Gas Type] > [O2] > [BACK] > [BACK].
(2). Correction Mode: Press the [Setup] button, and select [Setting] >
[ENTER] >[Correction Mode]> [BTPS]> [BACK]> [BACK]