8-18
Message
Alarm
Priority
Cause Solution
Mixed Agent Low
The AG module
supports measurement
and calculation of two
types of halogenated
anesthetic agent, and
the measured MAC is
smaller than 3.
Use only one type of halogenated anesthetic
agent.
Mixed Agent Medium
The AG module
supports measurement
and calculation of two
types of halogenated
anesthetic agent, and
the measured MAC is
invalid.
Use only one type of halogenated anesthetic
agent.
Mixed Agent
and MAC 3
Medium
The AG module
supports measurement
and calculation of two
types of halogenated
anesthetic agent, and
the measured MAC is
equal to or greater
than 3.
Use only one type of halogenated anesthetic
agent.
CO
2
Over
Range
Low
The monitored value
exceeds the
measurable range of
the module.
1. Reduce the concentration of the monitored
gas to the normal range.
2. Calibrate the AG module again.
3. Replace the AG module.
O2 Over
Range
Low
The monitored value
exceeds the
measurable range of
the module.
1. Reduce the concentration of the monitored
gas to the normal range.
2. Calibrate the AG module again.
3. Replace the AG module.
N2O Over
Range
Low
The monitored value
exceeds the
measurable range of
the module.
1. Reduce the concentration of the monitored
gas to the normal range.
2. Calibrate the AG module again.
3. Replace the AG module.
HAL Over
Range
Low
The monitored value
exceeds the
measurable range of
the module.
1. Reduce the concentration of the monitored
gas to the normal range.
2. Calibrate the AG module again.
3. Replace the AG module.
ENF Over
Range
Low
The monitored value
exceeds the
measurable range of
1. Reduce the concentration of the monitored
gas to the normal range.
2. Calibrate the AG module again.