8-21
Message
Alarm
Priority
Cause Solution
C High Imped. electrode C is too high. in proper contact with the skin.
2. Replace the BIS sensor.
BIS Electrode
C Lead Off
Low
BIS electrode C is
detached from the patient.
1. Check whether BIS electrode C is
in proper contact with the skin.
2. Replace the BIS sensor.
BIS Electrode
LE High
Imped.
Low
The impedance of BIS
electrode LE is too high.
1. Check whether BIS electrode LE
is in proper contact with the skin.
2. Replace the BIS sensor.
BIS Electrode
LE Lead Off
Low
BIS electrode LE is
detached from the patient.
1. Check whether BIS electrode LE
is in proper contact with the skin.
2. Replace the BIS sensor.
BIS Electrode
LT High
Imped.
Low
The impedance of BIS
electrode LT is too high.
1. Check whether BIS electrode LT
is in proper contact with the skin.
2. Replace the BIS sensor.
BIS Electrode
LT Lead Off
Low
BIS electrode LT is
detached from the patient.
1. Check whether BIS electrode LT
is in proper contact with the skin.
2. Replace the BIS sensor.
BIS Electrode
RE High
Imped.
Low
The impedance of BIS
electrode RE is too high.
1. Check whether BIS electrode RE
is in proper contact with the skin.
2. Replace the BIS sensor.
BIS Electrode
RE Lead Off
Low
BIS electrode RE is
detached from the patient.
1. Check whether BIS electrode RE
is in proper contact with the skin.
2. Replace the BIS sensor.
BIS Electrode
RT High
Imped.
Low
The impedance of BIS
electrode RT is too high.
1. Check whether BIS electrode RT
is in proper contact with the skin.
2. Replace the BIS sensor.
BIS Electrode
RT Lead Off
Low
BIS electrode RT is
detached from the patient.
1. Check whether BIS electrode RT
is in proper contact with the skin.
2. Replace the BIS sensor.
BIS DSC Err Low
BIS DSC communication
error
4. Replace the BIS module.
BIS DSC Malf Low
The BIS DSC is faulty
and is automatically shut
down.
4. Replace the BIS module.
BIS No Cable Low
The BIS cable is not
connected.
1. Check the BIS cable.
2. Replace the BIS cable.
BIS No Sensor Low
The BIS sensor is not
connected.
1. Check the BIS sensor.
2. Reinstall the BIS sensor.
BIS Sensor
Too Many
Low
The BIS sensor is
excessively used.
1. Check whether the use times of
the BIS sensor exceeds the limit

Table of Contents