In case of mechanical ventilation, during inspiration, gas flows through manual/auto switch 41, 
BYPASS valve 35 or sodalime canister 34, inspiratory valve 33, O2 sensor 39, airway pressure 
gauge 47, and inspiratory flow sensor 37 to the patient. During expiration, gas flows through 
expiratory flow sensor 38, expiratory valve 36 and manual/auto switch 41 to the folding bag or 
manual bag. Airway pressure is monitored by pressure sensor 53. 
 
The water generated by condensation is collected in the water collection cup on the bottom of 
breathing system. 
 
The breathing system is easily disassembled and is autoclavable at 134℃. 
43. APL valve 
41. Manual/auto 
switch 
45. Bellows 
assembly 
Bag arm 
47. Airway 
pressure gauge 
36. Expiratory 
valve 
33. Inspiratory 
valve 
Patient connection (built-in 
inspiratory and expiratory 
flow sensors 37 and 38) 
34. CO2 
absorbent canister 
35. Built-in 
BYPASS valve 
Handle