The correction for the interpretation of the data coming from the RH/T sensor (Calibration; %RH) - Menu
entry 11.
The hysteresis - Menu entry 12
Furthermore the following data can be read out from within the menu:
The temperature inside the case as read from the RH/T sensor (°C) - Menu entry 7.
The temperature on the condenser inside the EBC (°C) - Menu entry 8.
The temperature on the heat sink inside the EBC (°C) - Menu entry 9.
The hours of operation of the EBC (Hrs) - Menu entry 10.
IMPORTANT: Note down the current hours of operation of your EBC for future reference (cp. section 7.5
"General service" on p. 51).
NOTE: The menu is identical at masters and slaves, but some settings can only be taken on masters. If
you i.e. try to alter the setpoint or one of the alarm thresholds within the menu of a slave unit, the software
will switch back to the value set in the master shortly after.
NOTE: Any EBC will cancel the menu mode after a few seconds of inactivity. All changes that have been
made and then confirmed with ↵ Enter will remain intact. All changes that have not been confirmed before
will be rejected.
4.5.1 Menu entry Setpoint
With this entry you can set the desired level for the relative humidity inside your case (master units only) -
this is the core function of an EBC (or a chain of EBCs). Default setting is 50%RH (might also be factory-set
to a client-specific value). The EBC master unit always compares the value it receives from the RH/T sensor
inside your case to the one defined here. It then starts/continues with its work to bring the measured value
inside the case in line with the one set by you. All connected slaves will duplicate this work.
Press Menu/Esc to call up the menu of the master unit. The display reads:
S e t p o i n t 5 0 % ↵
Fig. 12 - Display message in menu mode, 1st menu entry.
Applies analogously to the individual products of series EBC10/11/12 (Rev.3). As of 24.11.2014. Subject
to modification.
miniClima - Manual_EBC10-11-12(Rev.3)_en
26 / 54