to pump water inside will fail and so the pump alarm will additionally be issued (see above).
NOTE: By default any EBC is equipped with one water level sensor for the high mark of the bottle. Some
accordingly ordered units also have a sensor for the low mark, and some have been ordered without any
means for monitoring the water level in the bottle.
5.3.2 Remaining EBC functions during this alarm
Humidify: Yes (until the remaining water in the bottle is used up)
Dehumidify: Yes, partly ( CAUTION: In case the max. level has been reached, mind to empty the bottle
in time! There will be enough time for this after the alarm has been issued. For safety reasons and in case
the alarm had been triggered by the max.-level sensor, the EBC will stop dehumidifying after four hours.)
Air circulation between case and EBC: Yes
Pump: Yes
Fan(s) for the device cooling: Yes
Slave control: Yes
5.3.3 Possible reasons and solutions
Please investigate in the outlined order:
1) Is the bottle either too full or is it becoming empty? -See section 4.7 "Controlling, feeding, emptying and
cleaning the bottle" on p. 37 on how to handle the bottle.
2) The bottle should actually touch the water level sensor in order to enable the sensor to issue the alarm.
The water level sensor is a turquoise rectangular plate mounted to the aluminium bottle holder, cp. Fig. 3
on p. 10). Use the black hook-and-loop-fastener for pulling the bottle towards the sensor for as much as
possible.
3) If the water level is in order, the surface of the bottle might have become soiled over the time. Verify the
function of your upper water level sensor by first pulling the bottle away from the sensor (alarm shall go
away with this) and by then putting a finger on the centre of the sensor. If the alarm comes back with your
finger on the upper sensor, cleaning the bottle should solve the matter. See section 4.7 "Controlling,
feeding, emptying and cleaning the bottle" on p. 37 on how to handle the bottle.
4) If neither of the above helps, do get in touch to discuss the next steps.
5.4 Signal error alarm
Applies analogously to the individual products of series EBC10/11/12 (Rev.3). As of 24.11.2014. Subject
to modification.