EasyManua.ls Logo

miniClima EBC - Bottle Maintenance; General Service Recommendations; Deinstalling, Packing, Storing, and Shipping the EBC

Default Icon
54 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
miniClima Schönbauer GmbH
Brunner Str. 21b, 2700 WIENER NEUSTADT, AUSTRIA, EU
Tel. | Phone: +43 2622 24964, +43 6991 24964 01, Voicemail: +43 6993 24964 01, Fax: +43 2622 24964 15, E-Mail: office@miniclima.com
www.miniclima.com, twitter.com/miniClima, www.linkedin.com/company/miniclima-schönbauer-gmbh, www.google.com/+Miniclima, www.facebook.com/miniclima
UID | VAT-ID: ATU61994435, EORI: ATEOS1000001229, DVR: 4008782
Firmenbuchnummer | Trade Register Number: 267137z (Landesgericht | Regional Court of Wiener Neustadt)
the mains plug before moving the device for inspecting the air slits.
Finally, do check regularly whether adequate ventilation is still given at the device's place of installation (see
section 3.2 "Setting up the EBC, establishing the hosepipe connections", p. 12).
7.4 Bottle
See section 4.7 "Controlling, feeding, emptying and cleaning the bottle" (p. 37).
7.5 General service
A general service of the device carried out by the manufacturer is recommended after 15,000-20,000 hours
of operation (equals approx. two years when permanently in use; see section 4.5.10 "Menu entry OpHours"
on p. 33 to see how to check for the hours of operation your EBC has been in use so far). Please contact us
to make an appointment (see page header). Please also read and follow the instructions on how to deinstall,
pack up, and store the device below.
8 DEINSTALLING / PACKING UP / STORING / SHIPPING
NOTE: If the device in question is to be shipped for service/repair, it is enough to send the EBC alone - all
other parts and installations (bottles, hosepipes, case-sided hosepipe fittings, RH/T sensors with their cables
and plugs, control lines, 6pin plug-in connectors, RS232 cables, Serial2USB adapters includ. cables, mains
connection cables, optional accessories) can stay in place.
1) Turn the EBC off (by pushing and holding On-Off/Reset for about a second). Slaves are switched
automatically with their master. Then, wait for the pump process to be completed (if any), so that no water
remains inside the EBC.
2) Pull the mains plug off the socket outlet, then pull the mains plug from the mains connection on the
backside of the device.
3) Loosen the hook-and-loop-fastener that fixes the bottle and pull the silicone pipe out of the bottle. Empty
and clean the bottle, if necessary. If the bottle is going to be stored over a longer period of time, tape the
small borehole on top of the bottle. IMPORTANT for transporting/packing devices with 2 litre bottles:
Reduce the risk of transport damages by mounting the bottle holder to some of the lower boreholes of the
anchor element on the front or side panel of the unit.
4) Pull all cables from the front plate of the EBC and dismantle the hosepipes on the backside. Models with
Applies analogously to the individual products of series EBC10/11/12 (Rev.3). As of 24.11.2014. Subject