1) Turn the EBC off (by pushing and holding On-Off/Reset for about a second). Slaves are switched
automatically with their master. Then, wait for the pump process to be completed (if any), and (
IMPORTANT) pull the mains plug.
2) Pull out the silicone pipe from the bottle and open the hook-and-loop-fastener.
3) Clean, feed or empty the bottle as required and close the bottle again. Mind that the bottle should always
be filled with at least some cm of distilled water so as to avoid the pump alarm being issued during
humidification. CAUTION: Do not spill water on the device! Danger of electric shock! Keep distance
from the device or any other, third party electric parts/installations when handling the bottle!
4) Put the bottle back on its place, secure it with the hook-and-loop-fastener (the bottle should touch the
sensor(s)), and push the silicone pipe inside again (the end of the pipe should to reach the bottom of the
bottle). Make sure to not having spilled water on the device, then push the mains plug in again.
5 ALARMS - WHAT THEY MEAN AND HOW TO REACT
miniClima Constant Humidity Devices are equipped with several alarms that go off when an intervention by
the user is needed. In the event of an alarm the EBC's LED turns red and the display provides immediate
information about the kind of problem. Additionally any alarm switches the potential-free contact for the
external display of the alarm status (see section 3.3.4 "Wirings for the external signals (Composite error
alarm and on-off status)" on p. 20. Pushing the On-Off/Reset button next to the LED shortly resets the alarms
according to the rules provided in Table 1 (p. 22).
5.1 Humidity alarm
5 0 % H u m i d i t y
5 0 % A L A R M
Fig. 24 - Display message for the humidity alarm. The digits highlighted in red blink. The words "Humidity"
and "ALARM" light up rotationally.
5.1.1 Description of the alarm
This alarm does not necessarily point to any sort of malfunction, but first of all provides the user with the
information that the desired humidity level has not (yet) been reached. The EBC will continue with its work
uninterrupted. The alarm appears when the actual RH value inside the case remains for at least 30 minutes
beyond one of the predefined thresholds (see sections 4.5.2 "Menu entry AlarmMax" on p. 27 and 4.5.3
Applies analogously to the individual products of series EBC10/11/12 (Rev.3). As of 24.11.2014. Subject
to modification.
miniClima - Manual_EBC10-11-12(Rev.3)_en
38 / 54