beneath your desired value and see whether the sought condition in the case will now be met.
6.2 The desired humidity level is not reached at all
Please read the comments on the humidity alarm (p. 38ff).
6.3 The humidity level displayed on the EBC is different to the one of a third party instrument that
has been put inside the case
Please do only use metered instruments to verify the humidity level measured by the EBC. Third party
instruments for measuring/verifying the humidity level might display different levels when positioned
differently (not exposed to the very same conditions). Also the accuracy of all used sensors (even metered
ones) has to be taken into account. Please, mind that measuring the relative humidity is never 100%
accurate. If you still feel the miniClima sensor gives out false values, consider using the correction menu on
the EBC front (see section 4.5.11 "Menu entry rH Corr" on page 34) or give us a call.
6.4 A red LED lights up on the front panel
The unit displays an alarm. Note the kind of the alarm (display messages) and turn to p. 38ff (section 5
"Alarms - what they mean and how to react") for further investigation.
6.5 When pulling or plugging in the sensor cable or the control line the EBC reacts in an unexpected
way
Never plug or unplug a sensor or control line during operation. Your EBC can become seriously damaged
through electrostatic discharges. Always turn all interconnected units off (via the master), wait for the
completion of the possibly following pump process and pull their mains connections before pulling or
plugging in those signal cables.
6.6 The bottle sensor for the high mark issues its alarm, despite the water level being far below the
sensor
This can happen when the inside of the bottle has been soiled over the time, so the sensor can't detect the
water anymore. See the possible reasons and solutions for the bottle alarm on p. 43.
Applies analogously to the individual products of series EBC10/11/12 (Rev.3). As of 24.11.2014. Subject
to modification.
miniClima - Manual_EBC10-11-12(Rev.3)_en