miniClima Schönbauer GmbH
Brunner Str. 21b, 2700 WIENER NEUSTADT, AUSTRIA, EU
Tel. | Phone: +43 2622 24964, +43 6991 24964 01, Voicemail: +43 6993 24964 01, Fax: +43 2622 24964 15, E-Mail: office@miniclima.com
www.miniclima.com, twitter.com/miniClima, www.linkedin.com/company/miniclima-schönbauer-gmbh, www.google.com/+Miniclima, www.facebook.com/miniclima
UID | VAT-ID: ATU61994435, EORI: ATEOS1000001229, DVR: 4008782
Firmenbuchnummer | Trade Register Number: 267137z (Landesgericht | Regional Court of Wiener Neustadt)
6.7 The bottle sensor for the low mark won't issue its alarm, despite the water level being far below
the sensor
This can happen when the inside of the bottle has been soiled over the time, so the sensor can't detect the
water anymore. Please check by pulling away the bottle from the sensor. If the bottle alarm is issued by
doing so, cleaning the bottle should solve the problem. See section 4.7 "Controlling, feeding, emptying and
cleaning the bottle" on p. 37 on how to handle the bottle. If the problem cannot be solved as described, get in
touch to discuss the next steps.
6.8 The device won't react in any sort of way anymore, neither the LED nor the display are on,
despite of the device being plugged on correctly
This behaviour points at a blown fuse. Other possible sources of error: A broken mains connection cable
(please verify by using a different (known to be working) cable with the affected EBC); a broken or non-
powered socket outlet (please verify by using a different (known to be working) socket outlet). In order to
replace the fuse, do please turn to section 7.2 "Replacing the miniature fuses" on p. 49. If neither of the
above applies and your EBC still won't react, the EBC needs to be sent in for inspection/repair (see section 8
"Deinstalling / packing up / storing / shipping" on p. 51).
7 MAINTENANCE AND CARE
7.1 Replacing the chip of the RH/T sensor
Even if it happens rarely, it might one day become necessary to replace the sensing part of the RH/T sensor
(hereinafter referred to as "chip")
12
. To do so, neither the housing, nor the cable, nor other parts need to be
dismantled, just the chip itself.
At first, please do turn the EBC off and pull the mains plug once the possible pump process has been
finished. Then continue as outlined:
1) Unscrew the plastic part of the housing:
12
A new chip can be ordered via your EBC supplier or directly with us.
Applies analogously to the individual products of series EBC10/11/12 (Rev.3). As of 24.11.2014. Subject
to modification.
miniClima - Manual_EBC10-11-12(Rev.3)_en
47 / 54