40 User Guide EN Dansensor® MAP Check 3 Vacuum
P/N 320598-G
11/2019
26 Error Import/Export: no data found User information
27 Error Import/Export: corrupted data User information
28 Error Import/Export failed User information
29 Warning CO
2
sensor needs calibration. Measuring values
may not be valid.
Contact service
for sensor
calibration
30 Warning O
2
sensor needs calibration. Measuring values may
not be valid.
Contact service
for sensor
calibration
31 Error Cannot set mixer gas ratios. Mixer is not configured
for gases found in active product.
User information
32 Error Unknown error from STM Contact service
33 Error Error from STM Contact service
34 Error Error from STM Contact service
35 Error Internal communication error Contact service
36 Error Internal communication error Contact service
37 Error External analog-digital converter failed Contact service
38 Error External analog-digital converter failed Contact service
39 Warning Atmospheric pressure sensor is not calibrated.
Pressure values may not be valid.
Contact service
40 Warning Difference pressure sensor is not calibrated.
Pressure values may not be valid.
Contact service
41 Error Voltage/current analog output failed Contact service
42 Error CO
2
sensor communication error Contact service
43 Error CO
2
sensor output out of range Contact service
44 Error Zr O
2
: could not detect cold-junction temperature
sensor
Contact service
45 Error Zr O
2
: error with heater regulator Contact service
46 Error Zr O
2
: cold-junction temperature sensor error Contact service
47 Error O
2
sensor temperature out of range Contact service
48* Warning Too high O2 sensor temperature: XX C Contact service
49* Warning Too low O2 sensor temperature: XX C Contact service
50 Error O
2
sensor is disabled because of hardware fail.
Please call service.
Contact service
51 Error O
2
sensor is disabled because of hardware fail.
Please call service.
Contact service
No. Type Message User action