Dansensor® MAP Check 3 Vacuum EN User Guide 41
P/N 320598-G
11/2019
52 Error O
2
sensor is disabled because of hardware fail.
Please call service.
Contact service
54 Warning Error in O
2
sensor calibration data. Check if the gas
concentration and span is correct.
Contact service
for sensor
calibration
55 Warning O
2
sensor needs calibration. Measuring values may
not be valid.
Contact service
for sensor
calibration
58 Error Fan speed too low, check if fan is functional Contact service
59 Error Error in fan parameters, running fan at full speed. Contact service
60 Error Temperature sensor not calibrated. Device
temperature may not be valid.
Check cooling/
filters
If error persists,
contact service
69 Error Sample system clogged. Check hose for blockages User information
70 Error Illegal mixer setting: a 2-gas mixer is connected but
3-gas mixer is expected
User information
71 Error Cannot establish connection to external mixer User information
72 Error Data log network dump failed: cannot connect to
server
Check LAN
connection
73 Error Data log network dump failed: connection to server
was lost
Check LAN
connection
74 Error Data log network dump failed: no ACK received Check LAN
connection
75 Error Data log network dump failed: wrong ACK received Check LAN
connection
76 Warning Mixer low pressure alarm on gas input X User information
77 Error Error from slave mixer User information
79 Error Zr O
2
cold-junction temperature too high. Possible
O
2
sensor fan fault.
Contact service
83 Warning Ambient pressure value out of reasonable range.
Possible pressure sensor fault.
Check and clean
Air inlet/Sensor
gas/sample gas
outlets
If error persists,
contact service
No. Type Message User action