Dansensor® MAP Check 3 Vacuum EN User Guide 43
P/N 320598-G
11/2019
92 Error Ir CO2 sensor temperature too low, shutting down
CO2 sensor.
Check ambient
temperature -
must be within
range
If problem
persists, contact
service
93* Warning Ir CO2 sensor temperature high Check ambient
temperature -
must be within
range
Check cooling/
filters
If problem
persists, contact
service
94 Error Ir CO2 sensor temperature too high, shutting down
CO2 sensor.
Check ambient
temperature -
must be within
range
Check cooling/
filters
If problem
persists, contact
service
98 Error Input pressure too high Reduce pressure
at gas input
99 Error Input pressure too low Increase pressure
at gas input
101 Warning Product uses external mixer but mixer is disabled Enable mixer or
choose different
product
102 Warning Buffer alarm Check alarm
settings for
buffer.
103 Event TIMING PROBLEMS, VAC signal too soon Check control
signals
104 Event TIMING PROBLEMS, IR delay too long Check control
signals
105 Error Critical high pressure on buffer input. More than 10
bar may be destructive
Lower input
pressure on
buffer input
232 Error Vacuum is above max. Vacuum=%s, Max=%s Check vacuum
No. Type Message User action