117
Technical Specications
Down Occlusion Detection
Time
Microbore tubing with
occlusion at distal Luer
For non-ltered sets:
Rate
0.1 mL/hr
1 mL/hr
10 mL/hr
50 mL/hr
125 mL/hr
Low Pressure
≤ 1 hr. 30 minutes
≤ 7 minutes
≤ 40 seconds
≤ 6 seconds
≤ 3 seconds
High Pressure
≤ 2 hrs. 15 minutes
≤ 16 minutes
≤ 1min. 30 seconds
≤ 15 seconds
≤ 5 seconds
For ltered sets:
Rate
0.1 mL/hr
1 mL/hr
10 mL/hr
50 mL/hr
125 mL/hr
Low Pressure
≤ 9 hr. 15 minutes
≤ 50 minutes
≤ 4 min. 30 seconds
≤ 50 seconds
≤ 20 seconds
High Pressure
≤ 20 hrs.30 minutes
≤ 2 hrs. 5 minutes
≤ 10 min. 30 seconds
≤ 3 minutes
≤ 50 seconds
ITEM SPECIFICATION
Programmable Time (HH:MM)
PCA
No time limit
Bolus Volumes
Microbore tubing with
occlusion at
distal Luer
Low Pressure: <0.1 mL
High Pressure: <0.1 mL