EasyManua.ls Logo

Moog CURLIN INFUSION PainSmart - Page 136

Moog CURLIN INFUSION PainSmart
141 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
129
1. The setting is changed by using either the “BioMed SETUP” function or the OPTIONS
function.
Note: Entering a new program will NOT change the Lock Level from its current setting.
2. The user selects a preprogrammed Protocol from a Protocol Library Safety System-
enabled pump. In this case, the pump defaults to the Protocol and erases the Lock Level
setting.
If a locked-out function is requested, such as PRIME in lock setting 3, a screen will appear
requesting the entry of the access code. When the correct code is entered, the locked out function
can be performed by the clinician.
Note: In any function outside of the field to change the Lock Level, entering the clinician access
code will not change the Lock Level setting. Once the pump recognizes the clinician access code, it
authorizes the function to be performed without the need to change the Lock Level setting.
Clinician Information

Table of Contents

Related product manuals