29
e battery pack provides 5.2V with a capacity of 6000 mA-hrs. It is connected to the same 
port connector that the AC adaptor uses to power the pump. When the battery pack power is 
fully exhausted the pump will automatically revert to the internal C-cell batteries. Having new 
C-cell batteries in the pump is crucial to ensure an adequate low battery warning period and to 
ensure infusion continuity if the battery pack is accidently disconnected. e red Low Battery 
LED on the battery pack will illuminate prior to the external battery running out of power. 
e pump will provide at least 30 minutes of audible and visual warning prior to the internal 
C-cell batteries running out power. 
e battery pack is recharged by connecting it to an AC adaptor. Fully recharging the battery 
takes approximately 15 hours. While charging, the LED indicator will be yellow. When the 
battery is fully charged, the LED will be green. If a fault is encountered in the battery pack the 
LED will either ash yellow or not illuminate.
 
Warnings
•  The external battery pack should be used with new C-cell batteries in the pump to avoid 
interruption of the infusion if the cord between the pump and battery pack becomes 
disconnected and to ensure proper operation of the pump’s low battery alarm. 
•  Do not use this battery pack with the Moog Curlin 340-2022 (POWDEC AC Adaptor 
WM10-6.9V)   as the combination may exceed the allowable limits for emitted electromagnetic 
interference.
•  Battery pack cable utilizes an attached cable. This can be a tripping or chocking hazard, avoid 
use that may lead to entanglement around the neck or feet. 
Cautions
•  This battery pack is fluid resistant, but it is not designed to be water tight. Care should be 
taken to avoid any unnecessary contact with water or other fluids. Do not use the battery pack 
where it might fall into water. 
•  The power gauge located on the pump screen (pictured below) does not reflect the remaining 
charge in the battery pack. The displayed capacity on the pump will always show empty when 
a battery pack (with remaining capacity) is attached (see image below): 
Pump & Set Features