HYDROCAL 1001+ Installation and operation manual Page 54/76
I - Test Gas concentration
Step
Description
Current
[mA]
Start
-
Red on / green on
actual value
1 x
1 of 6
Red on / green off
0
1 x
5 x
2 of 6
Red off / green on
4
1 x
5 x
3 of 6
Red off / green ~1 s on / ~1 s off
8
1 x
5 x
4 of 6
Red off / green ~½ s on / ~½ s off
12
1 x
5 x
5 of 6
Red off / green ~¼ s on / ~¼ s off
16
1 x
5 x
6 of 6
Red off / green ~ s on / ~ s off
20
1 x
5 x
a
II - Test Moisture concentration
Step
Description
Current
[mA]
a
-
Red on / green on
actual value
1 x
1 of 6
Red on / green off
0
1 x
5 x
2 of 6
Red off / green on
4
1 x
5 x
3 of 6
Red off / green ~1 s on / ~1 s off
8
1 x
5 x
4 of 6
Red off / green ~½ s on / ~½ s off
12
1 x
5 x
5 of 6
Red off / green ~¼ s on / ~¼ s off
16
1 x
5 x
6 of 6
Red off / green ~ s on / ~ s off
20
1 x
5 x
b
III - Test Function alarm
Step
Description
Jumper J1 on Pos. 1 (J1:1-2)
Relay-
contact
b’
-
White color / contact closed
X2:7/8
1 x
1 of 2
Red color / contact open
X2:7/8
1 x
5 x
2 of 2
Green color / contact closed
X2:7/8
1 x
5 x
c
Step
Description
Jumper J1 on Pos. 1 (J1:2-3)
Relay-
contact
b’’
-
White color / contact open
X2:7/8
1 x
1 of 2
Red color / contact closed
X2:7/8
1 x
5 x
2 of 2
Green color / contact open
X2:7/8
1 x
5 x
c

Table of Contents