EasyManua.ls Logo

Newport 2832-C - Page 66

Newport 2832-C
118 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
CH?
Available Channel
Query
Syntax:
CH?
Parameters:
None
Function:
This query
is
used to see which channels are available in the meter.
A
chan-
nel
is
available
if
a calibration module was plugged into that channel on
powerup and passed the self test.
Returns: <channels>
<channels>
is
of type <string> in the double quote format. The possible values
of <string> are:
"A"
-
only channel
A
is
available
"B" -only channel B
is
available
"AB
-
Both channel A and B are available
Related Commands: None
DnMODELn?
Detector Model
Query
Syntax:
DETMODEL-n?
where n
is
either
A
or B
Parameters:
None
Function:
This query returns the model number of the detector that belongs to the
memory module attached to the specified channel.
Returns:
<model>
<model>
is
of type <string> in the double quote format.
Related Commands: A7TNSN-n?, CALDATE-n?, DETSN-n?
DEISN-n?
Detector
Serial
Number Query
Syntax:
DETSN-n?
where n is either A or B
Parameter:
None
Function:
This query returns the serial number of the detector that belongs to the
memory module attached to the specified channel.
This serial number must match the serial number of the detector attached to
the
28324.
Returns: <detSN>
<detSN>
is
of type <string> in the double quote format.
Related Commands: ATTNSN-n?, CALDATE-n?, DETMODEL-n?
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Table of Contents

Related product manuals