TRIGOUTLVL
Trigger Out Level Select
Syntax:
TRIGOUTLVL
<level>
Parameters:
The parameter <level>
is
of type <number>. The units
of
this parameter are
Watts.
Function:
This command
is
used to set the level used by the trigger out level compara-
tor mode. When the power level is at or above this level the trigger out will go
to its active state.
Related Commands:
TRIGOUT, TRIGOUT?, TRIGOUTFREQ, TRIGOUTFREQ?,
TRIGOUTLVL?, TRIGOUTPOL, TRIGOUTPOL?
TRIGOUTLVL?
Trigger Out Level Query
Syntax:
TRIGOUTLVL?
Parameters:
None
Function:
This query returns the trigger out level used by the trigger out level
comparator mode.
Returns:
<level>
<level>
is
of type <number> in exponent notation with units of Watts.
Related Commands:
TRIGOUT, TRIGOUT?, TRIGOUTFREQ, TRIGOUTFREQ?,
TRIGOUTLVL, TRIGOUTPOL, TRIGOUTPOL?
TRIGOUTPOL
Trigger Out Polarity Select
Syntax:
TRIGOUTPOL
<state>
Parameters:
The parameter <state> is a <number> which can be
0
or
1.
If
the number
rounds to
0,
then the polarity will be active low.
If
the number is
1,
then the
polarity will be active high.
Function:
This command
is
used to set the polarity of the trigger out output to either
active low or active high.
Related Commands:
TRIGOUT, TRIGOUT?, TRIGOUTFREQ, TRIGOUTFREQ?,
TRIGOUTLVL, TRIGOUTLVL?, TRIGOUTPOL?