Training Guide Course No. 196514
Level 2 Maintenance, X-1000 Series Dispensing Systems 4-121 P/N 196515 (Revision A)
Is mechanical
correction required as
a result of the above
inspections?
Resolve and go to Scale
Operational Verification
Go to Scale Operational
Verification
• Scale Control Module and the Balance Module have matching serial numbers.
• Scale has a valid date on the calibration label.
• Scale Pedestal not damaged and has no residual fluid build-up.
• Scale Cup Cover properly in place.
• Scale Balance properly installed and leveled.
• No loose, damaged, or pinched cables.
• No noticeable vibrations in the dispensing system or adjacent machinery
• System installation properly leveled and footpads firmly seated on floor.
Inspect and confirm that the following conditions are present:
Yes No
Is mechanical
correction required as
a result of the above
inspections?
Is mechanical
correction required as
a result of the above
inspections?
Resolve and go to Scale
Operational Verification
Go to Scale Operational
Verification
• Scale Control Module and the Balance Module have matching serial numbers.
• Scale has a valid date on the calibration label.
• Scale Pedestal not damaged and has no residual fluid build-up.
• Scale Cup Cover properly in place.
• Scale Balance properly installed and leveled.
• No loose, damaged, or pinched cables.
• No noticeable vibrations in the dispensing system or adjacent machinery
• System installation properly leveled and footpads firmly seated on floor.
Inspect and confirm that the following conditions are present:
• Scale Control Module and the Balance Module have matching serial numbers.
• Scale has a valid date on the calibration label.
• Scale Pedestal not damaged and has no residual fluid build-up.
• Scale Cup Cover properly in place.
• Scale Balance properly installed and leveled.
• No loose, damaged, or pinched cables.
• No noticeable vibrations in the dispensing system or adjacent machinery
• System installation properly leveled and footpads firmly seated on floor.
Inspect and confirm that the following conditions are present:
Yes No
Diagram Sheet 4-9-1
Scale Fault Isolation Procedure – Mechanical Verification