5-3
5. Linearity
Instructions for Use Manual 5 Linearity Test
8. With the test strip correctly inserted, the Apply
Sample screen displays.
9. Gently mix the StatStrip Linearity Solution before
each use.
10.Discardtherstdropoflinearitysolutionfromthebottle
to avoid contamination.
11. Place a drop of linearity solution from the bottle at the
end of the test strip until the solution is drawn into the
well of the test strip. When enough sample has been
drawn into the strip, an audible beep is sounded by
the meter.
Figure 5.4 Apply Sample (Linearity Solution) to Test Strip Screen
13:42
Touch test strip to Level 1 Lin drop
Remove test strip to cancel test
Op: ****
Name: A Unit-1
StatStrip Glu IFU 1.86 PN55848E EN 510K.indd 3 9/13/18 8:29 AM