iCare+ Operation Manual 10-8-1 p. 26
ENGLISH
In parallel of the function of external spray instrument, the unit can control if there is a mistake for the programming
of the button SPRAY or if a channel is blocked.
For example, if the user selects an instrument as an external spray instrument (but in reality, the instrument has an
internal spray), the unit is able to detect it.
As a reverse case, if the user connects an instrument with external spray without using the button “ext W/O” (to
select external spray: “ext-w/o” LED), the unit is able to detect that the selection is wrong because this instrument
has external spray.
To inform the user of these problems, the unit will stop and the Air LED will blink, with a BIP signal heard. In
parallel, the concerning instrument is indicated by white LED blinking (oil indicator and/or ext-WO LEDs).
To restart the unit, first open the door, remove the instrument (to reduce the remaining pressure) reconnect the
instrument, and then close the door.
If the problem happens again, remove the handpieces and push the button “Start”.
Check the instruments (in case of possible blocked channels).