iCare+ Operation Manual 10-8-1 p. 38
ENGLISH
For example, if the user selects an instrument as an external spray (but in reality, the instrument has an internal
spray), the unit is able to detect it.
As a reverse case, if the user connects an instrument with external spray without using the button SPRAY (to select
external spray: ext-w/o” LED), the unit is able to detect that the selection is wrong because this instrument has
external spray.
Whatever the reason of the alert is (blocked channel or wrong selection), to restart the unit, it is first necessary to
check the reason of the problem. Then, open the door, disconnect the instruments and reconnect them. Finish by
closing the door.
Once this is done, push “START” button to reset the error. The Air LED should be switch off.
Check the instrument, if one of the channels is not blocked, and program the right selection (internal or external
spray instrument) by using the ext-WO button + the number of the instrument.
8.1.5. Door sensor
Corresponding red LED “Door” is lit (No “BIP signal).
Please, make sure that the door is properly closed.
8.2. Fatal error
The 6 red LEDs are blinking and 3 “BIP” sounds are heard.
If this case happens, please contact your retailer directly.