iCare+ Operation Manual 10-8-1 p. 18
3.3 Validation test
3.3.1. On site validation test
This test must be performed prior to first start-up. It consists of a series of tests in order to validate that the
functioning parameters are functioning after the installation of iCare+.
3.3.2. Periodical validation test
NSK recommend to perform this periodical validation test every year, in order to validate that the performance of
iCare+ is still optimum.
3.4. Connection of the instruments
3.4.1. Treated instruments
The iCare+ is a device whose function is to ensure the cleaning and disinfection of dental rotary and dynamic
instruments.
The iCare+ device allows the cleaning and disinfection of internal and external surfaces of different instruments. The
main instruments intended to be treated are handpieces: contra-angles and turbines.
The different types of instruments that can be treated by iCare+ device are listed in the table 1 (please refer to the
part 3.3.5. Selecting the type of instrument).
3.4.2. Installation of the instruments
With the iCare+ device, it is possible to connect 2 kinds of instruments by means of 2 different connectors:
Air turbine connector for turbines.
E-type connector for contra-angles.
To connect a turbine with Midwest 4 holes(in accordance with the ISO 9168 standard) , start by opening the door
and connect the turbine directly to turbine connector of the iCare+ (see Figure 14) and screw the ring on tightly (see
Figure 15). Finish this operation by closing the door. Take care of the position of the turbine connector.
Do not use the door as a support : do not put anything on the door when open.