iCare+ Operation Manual 10-8-1 p. 13
ENGLISH
CHAPTER 3: USE of iCare+ DEVICE
CAUTION: The use of controls, settings or the implementation of procedures other
than those specified below may expose users and patients to danger
3.1. Setup
3.1.1. Installing the iCare+ device
You should install the iCare+ device as requested below:
The device should be positioned on a flat, solid and level surface. The
maximum weight of the device, fully loaded, in conditions of use is 16 kg
The device should not be placed near a sink or any source that may splash it
The device should be placed in a room properly ventilated
The device should be kept away from any source of heat
3.1.2. Connecting the air supply
Insert the air supply tube in the air filter: J, fixed on the Rear panel's device. Be sure that
the air tube is attached properly to the filter (see Figure 6). Connect the "T-shaped fast-on"
connector to the surgery compressor's air supply.
Following the norm EN ISO 7494-2, the air quality provided to iCare+ has to
be dry, clean and free from bacteria and contamination.
The pressure of the air supply should be between 5 and 6 bar
The minimum air flow level provided to the unit is 50 L./mn.
It is necessary to follow the above mentioned parameters in order to avoid
some dysfunctions or damage of the unit.
Figure 6: Connection of the tube in the
pneumatic system of iCare+ device