In order to facilitate the external cleaning, we recommend using disinfectant
wipes (without protein fixation)for handling the instruments while
connecting them to the unit and rub slightly them before connecting in case
of important soils. It will help to liquefy the soils and stains located on the
outside surfaces.
At the end of the cycle, open the door and remove the instrument with a
disinfectant wipe (without protein fixation).
CLEANING PHASE (X3) :
- Injection of n. Clean product inside the
instrument (W/A + body channels)
- Simultaneous injection of n.Clean
product outside the instruments
(external surfaces)
- Action time of the liquid (30 s)
DISINFECTION PHASE (X1) :
- Injection of n.Cid product inside the
instrument (W/A + body channels)
- SImultaneous injection of n.Cid
outsite the instruments (external
surfaces)
- Action time of the liquid (120 s)
- Air blowing time
LUBRICATION PHASE (X1) :
- Injection of Oil inside the
instrument (body of the
instrument)
- Air blowing time