This symbol is designed to catch the reader's attention to any text containing vital
safety information. You are asked to read closely and understand properly each of the
sections concerned immediately before using your iCare+ device
or even accident.
Use the iCare+ device only indoors
Do not install or use iCare+ device near a naked flame owing to the risk of
explosion
Do not install or use iCare+ device in direct contact with sunlight
Ensure that the iCare + device is placed in a ventilated place
Store iCare+ device at temperature between 0°C and 50°C and humidity rate
under 80%
Use iCare+ device in a temperate room (22-40°C)
Do not place n.Cid bottles near a heat source
Do not turn the device upside down
Place the device horizontally on a flat surface
Only use NSK products when operating iCare+ device (n.Clean as cleaning
product, n.Cid as disinfection product, NSK Oil as lubrication product)
During the installation of iCare+ device, leave a space of 5 cm free on each
side
Be sure that air supply pressure provided by iCare+ device is set between 5
and 6 bar
Do not remove the bottles (n.Clean or n.Cid) while iCare+ device is in use
If not using iCare+ device for a long period, switch of it and cut the air arrival
Only use NSK components for the maintenance of iCare+ device. The use of
different components may damage the device