11 Maintenance parts
Light source lamps
Lamp name Model type Average lifetime
Halogen bulb
12V100WHAL-L (long lifetime type)
(made by PHILIPS Co. 7724I)
2,000 h
12V100WHAL (high light intensity type)
(made by PHILIPS Co. 7023)
100 h
Mercury burner
USH-103OL
(made by Ushio Co.)
300 h
Mercury burner for light
guide light source
SHI-130OL
(made by Ushio Co.)
2000 h
Halogen bulb for light guide
light source
JCR12V-100WB
(made by Ushio Co.)
1000 h
Filter
Filter name Model type Average lifetime
Reflected light band-pass
filter for IR
U-BP1100IR / U-BP1200IR
(made by Olympus)
500 h
Immersion oil
Olympus immersion oil IMMOIL-F30CC
· Olympus immersion oil : IMMOIL-F30CC
Obtain the Safety Data Sheet from the Website of Olympus Corporation.
· Absolute alcohol* (commercially-available product)
· Neutral detergent* (commercially-available product)
* Obtain the Safety Data Sheet from the supplier where you purchased this commercially-available product.
11-1 Maintenance parts list
11-2 List of chemicals used for maintenance