Important Information — Please Read Before Use
VISERA XENON LIGHT SOURCE CLV-S40
Xenon light sources produce significant heat due to the high
intensity light required for endoscopic procedures. If the
disconnected end of the light guide cable or the distal end of
an endoscope touches operating room drapes or other
flammable materials, fire can result.
Never place the unattached end of an illuminated light
guide cable on an operating room drape or other
flammable material.
Never allow the distal end of an illuminated endoscope to
come in contact with operating room drapes or other
flammable material.
Be sure to prepare another light source to avoid that the
examination must be interrupted due to equipment failure or
malfunction.
This product may interfere with other medical electronic
equipment used in combination with it. Before use, fully
confirm the compatibility of this instrument with all equipment
to be used with it.
Do not use this product in any place where it may be subject
to strong electromagnetic radiation (for example, in the
vicinity of a microwave therapeutic device, MRI, wireless set,
short-wave therapeutic device, cellular/portable phone, etc.).
This may impair the performance of the product.
Do not remove the cover of the light source. Electrical shock
can occur.
Do not use a pointed or hard object to press the switches on
the front panel of the instrument. This may break the
switches.
To avoid damaging electrical contacts and/or causing a
malfunction, do not apply excessive force to any connector.