42
Chapter 4 Inspection
VISERA XENON LIGHT SOURCE CLV-S40
If the high intensity mode indicator does not go off after the
endoscope is disconnected, immediately stop using the light
source and contact Olympus.
• The following endoscopes and light guide cables are
compatible with the high intensity mode:
− Light guide cable
A3290/91/92/93/94/95/96/97/98
− Fiberscope
CHF TYPE CB30L/S,
URF TYPE P3
− Videoscope
LTF TYPE V/V2/V3,
A4941A/43A
A4801A/03A/05A
A50001A/03A/21A/23A
• This instrument includes a memory function for the intensity
mode, which stores the intensity setting being used when the
light source is turned OFF. Therefore, the user may start the
next operation with the same setting used when the light
source was turned OFF.
• The default setting of this function is normal intensity.
4.7 Inspection of brightness adjustment
This function can be inspected when the brightness control indicator is lit.
Switching the brightness control (automatic/manual)
• When this instrument and the VISERA video system center
OTV-S7V are combined with any of the following products,
the brightness control indicator and MANUAL indicator are
disabled, and it is not possible to control the brightness from
this instrument. In this case, use the exposure level switch on
the OTV-S7V.
Refer to the instruction manual of the OTV-S7V.
− OTV-S7V camera head
− LTF TYPE V3
− A50001A/03A/21A/23A