Chapter 6 Lamp Replacement
59
VISERA XENON LIGHT SOURCE CLV-S40
Chapter 6 Lamp Replacement
Do not touch the inside of the lamp access cover or the lamp
immediately after use. These will be extremely hot and can
cause burns.
Disconnect the power cord from the wall mains outlet before
lamp replacement. Otherwise, electric shock can result.
When replacing the lamp, do not leave any objects (such as
a lint-free cloth) inside the lamp access cover. Fire and/or
equipment damage can result.
Never use a lamp that has not been approved by Olympus.
The use of a non-approved lamp can cause damage to the
light source and ancillary equipment, malfunction and/or fire.
Handle the lamps carefully in order to avoid the risk of
explosion.
Do not apply shock, excessive force or scratches to the lamp.
It could break the glass and/or shorten the lamp life due to
the high internal pressure of the lamp.
When disposing of the used main lamp, make sure to cut the
projection part on the lamp’s “+” side by using a nipper and
discharge the internal gas (see Figure 6.1). Otherwise, the
glass could break due to the high internal pressure of the
lamp.
When cutting the projection part on the lamp’s “+” side, wear
appropriate protection equipment all the time and be sure to
cover the lamp with cloth. Otherwise, the projection part
could scatter in all direction and cause injury to people
around. The protective equipment may include eye wear,
face mask and gloves.